top of page

CCM-iBIS

CCM-iBIS

Conveniently placed circular arm openings, a large front window and two side doors provide easy access for daily operations such as cleaning and disinfection. CO2 gas is supplied through the gas port located underneath the chamber, and gas concentration can be set from 0.0 to 10.0%.

Specification


Main Body

External dimensions

(W)382mm x (D)590mm x (H)219mm* (Projections excluded)

Number of settable dishes

9 (25-Well EmbryoGrid)

Inside-chamber dimensions

62 mm (W) x 65 mm (D) x 17 mm (H) Capacity: Approx. 69 ml

Humidification method

Non humidification

Power supply

220-240 VAC 50/60Hz 1.5A (Voltage fluctuation : within ±10%)

Weight of the main body

Approx. 30kg

Temperature control

External dimensions

(W)382mm x (D)590mm x (H)219mm* (Projections excluded)

Temperature control method

Digital PID method, independent control of each incubation chamber

Operational temperature range

24ºC to 28ºC

Temperature control range

36ºC to 38ºC

Temperature fluctuation range

±0.1ºC (Room temperature: 26ºC)

Temperature distribution

±0.1ºC (Room temperature: 26ºC)

Main heater

Silicon rubber heater

Gas Specifications

CO2 sensor

Infrared CO2 sensor

CO2 control method

Duty control

CO2 control range

0 to 20.0%

CO2 fluctuation range

±0.1%

O2 sensor

Ceramic O2 sensor

O2 control method

Duty control

O2 control range

2.0 to 18.0%

O2 fluctuation range

±0.5%

XY stage

Shift range

±105 mm (x-axis), ±140 mm (y-axis)

Resolution

1μm/step

Maximum speed

40mm/sec

Repetitive-positioning accuracy

Within ±0.5 μm

Z stage

Shift range

±5mm

Resolution

1 μm/step (Half)

Maximum speed

10mm/sec

Repetitive-positioning accuracy

Within ±0.2 μm

Optical System

Camera

1.3-million pixel CCD camera, 4.86 mm x 3.62 mm imaging size

Objective lens

Lens (10 x) for biological microscopes

Light source for illumination

Red LED (Peak wavelength: 623 nm)

Image

Image save format

JPEG format

Image save method

NAS (Mirroring function-equipped, 6TB x 2)s

External output

Analog output

Temperature (9 incubation chambers), CO2 concentration, O2 concentration

Contact output

Close when an alarm is generated; close when the power is turned Off


Copyright © 2025 Deworks Sdn Bhd.  All rights reserved.